Case Report

Volume 2, Number 5, October 2011, pages 201-205


Inferior Vena Cava Filter Migration to the Right Ventricle: A Case Report and Review of Filter Migration and Misdeployment

Jason Dreyera, d, Ketul Patelb, Huma Shujaatb, Michael Leec

aDepartment of Neurosurgery, Providence Hospital and Medical Center, Southfield, MI, USA
bDepartment of Internal Medicine, Providence Hospital and Medical Center, Southfield, MI, USA
cDepartment of Cardiothoracic Surgery, Providence Hospital and Medical Center, Southfield, MI, USA
dCorresponding author: Dr. Jason Dreyer, Providence Hospital, 16001 W 9 Mile Rd. Southfield, MI 48076, USA

Manuscript accepted for publication June 22, 2011
Short title: IVC Filter Migration
doi: https://doi.org/10.4021/jmc240w

Abstract▴Top 

Inferior vena cava (IVC) filters are often seen as a safe and effective means of preventing pulmonary embolus in patients with contraindication to standard medical therapy. We present a case of a 57-year-old African-American man who presented with an IVC filter in his right ventricle. At our institution, percutaneous retrieval is initially attempted to reduce perioperitive morbidity.

Keywords: Inferior vena cava; IVC; Filter; Migration; Misdeployment; Complication

Introduction▴Top 

Venous thromboembolism (VTE) is a significant cause of morbidity and mortality worldwide. The incidence of VTE is approximately 100 individuals per 100,000 each year. Deep venous thrombosis (DVT) accounts for approximately 66% of all VTE [1]. The vast majority of VTE are effectively managed with medical therapy such as systemic anticoagulation. Thrombi formation in the lower extremities or pelvis can embolize proximally to the pulmonary arteries. For this reason, inferior vena cava (IVC) filters are often utilized as a safe and effective means of preventing pulmonary embolism (PE). Interruption of the inferior vena cava has been employed in the form of IVC filters in a subset of patients in whom anticoagulation is not suitable. The subset would include persons who are at high risk for developing PE, who are unresponsive to anticoagulation therapy, or in whom anticoagulation is contraindicated (i.e. recent neurosurgery, craniospinal trauma, hemorrhagic stroke, gastrointestinal bleeding, recurrent DVT or PE despite anticoagulation).

In the largest randomized study of patients with symptomatic DVTs, IVC filter placement along with anticoagulation reduced incidence of developing PEs versus anticoagulation alone (1% and 5% respectively). This benefit was seen during the first 12 days after randomization; however no statistical difference was seen in patients at two years with respect to survival or symptomatic PE. At eight years, overall mortality was similar in both groups [2].

There are a variety of IVC filters available, each with its own characteristics, advantages, and disadvantages, which have been proven effective at preventing PE. Currently there are scarce reports of adverse events related to the Cordis TrapEase filter in the form of migration [3-5]. At our institution the TrapEase filter is favored because of its relatively simple placement, low profile introducing system, and low rate of complications.

We describe a case of a man who presented 8 days after placement of a Cordis TrapEase IVC filter, which migrated to the right ventricle and was subsequently surgically removed.

Case Report▴Top 

A 57-year-old African-American man with mild mental retardation, epilepsy, hypertension, diabetes mellitus, and chronic kidney disease, presented to the emergency department with complaints of sudden-onset shortness of breath, chest pain and “heaviness,” associated with profuse sweating. Further workup was performed and the patient was diagnosed with acute DVT of the left lower extremity, and a pulmonary embolus of the right upper lobe.

An extensive discussion regarding the risks and benefits of anticoagulation was performed. Due to his seizure disorder and recent history of falls of unidentified cause (5-6 in the last month), the decision was made to place an IVC filter.

The patent was initially placed on a heparin drip and interventional radiology was consulted. A venacavogram was performed and he was found to have a normal caliber IVC (Fig. 1A). A Cordis TrapEase IVC filter was deployed at the origin of the right renal vein without complication (Fig. 1B). The patient tolerated the procedure well, and he was discharged the following day.

Figure 1.
Click for large image
Figure 1. (A) Inferior vena cavogram showing a diameter of approximately 21 mm. (B) Deployed TrapEase filter (between arrows).

The patient was brought to the Emergency Department eight days later by his mother. She stated that “he is sleeping all day,” which was uncharacteristic for him.

Chest radiography revealed a foreign body consistent with an IVC filter in the vicinity of the tricuspid valve (Fig. 2). Echocardiography demonstrated an echo-refractile structure within the right ventricle (RV) with multiple echo-densities within the right atrium (RA), RV, and right ventricular outflow tract (RVOT) (Fig. 3). The laboratory values were significant for a platelet count of 34,000/µL. The patient was anticoagulated with argatroban, a direct thrombin inhibitor, pending the results of a heparin-induced thrombocytopenia (HIT) panel. Thrombotic thrombocytopenic purpura (TTP), idiopathic thrombocytopenic purpura (ITP), and HIT were considered and ruled out following a negative peripheral smear and the HIT panel.

Figure 2.
Click for large image
Figure 2. Lateral chest radiograph showing the location of the IVC filter in the heart (arrow).

Figure 3.
Click for large image
Figure 3. Apical, four-chamber, echocardiogram showing the presence of an IVC filter in right ventricle (arrow). The right atrium (RA), left ventricle (LV), and left atrium (LA) are labeled.

Interventional radiology determined that percutaneous retrieval of the IVC filter was too high risk due to the involvement of the chordae tendineae. Therefore, the patient was admitted to the cardiovascular intensive care unit for urgent cardiothoracic evaluation and treatment.

The patient underwent exploratory cardiac surgery. Utilizing cardio-pulmonary bypass, an excision of the 3.5 ×3.5 × 5.0 cm Cordis TrapEase IVC filter from the RV was performed. Multiple cords of thrombi were removed, and the damaged tricuspid valve was repaired (Fig. 4, 5).

Figure 4.
Click for large image
Figure 4. Excision of the IVC filter from the right ventricle (arrow). The right atrium (RA) is labeled.

Figure 5.
Click for large image
Figure 5. Excised IVC filter (arrow) with multiple cords of thrombi.

The patient’s post operative course was complicated with neurological deterioration requiring further investigations. Magnetic resonance imaging (MRI) of the brain demonstrated findings most consistent with bilateral watershed zones of infarction, likely related to a hypotensive/low flow state. Electroencephalogram demonstrated a diffusely encephalopathic state. Due to the patient’s prior comorbidities and multiple organ failure, the decision was made to place the patient under the care of hospice on hospital day fourteen. At the request of the family, he was taken to home hospice, and died three days later.

Discussion▴Top 

IVC filter placement to achieve partial mechanical interruption of the inferior vena cava is a well established method of preventing life-threatening PE caused by DVT [6]. In 1865, Armand Trousseau proposed the concept of creating a physical barrier to prevent emboli. Over the years many advances have been made to the concept of vena caval interruption. It initially began as surgical ligation of the femoral veins in the 1930s, and eventually led to the development of the Mobin-Uddin filter in 1970. The stainless-steel Kimray-Greenfield filter, introduced in 1973, revolutionized the field of vena caval interruption. In the 1980s further advancements made it possible for percutaneous placement of the IVC filters, and more recently, the development of retrievable filters [1].

The many styles of IVC filters on the market can be broken into three general categories: permanent, optional, and temporary. Permanent filters are intended to be left indefinitely within the IVC. Optional filters may be left indefinitely, but have a retrieval system if necessary. Temporary filters are intended to be removed after a short period of time, and remain attached to the patient’s skin by an intraluminal fixation system.

Each IVC filter type and style has its own advantages and disadvantages. Characteristics such as available access site, profile of introducer, patency of the filter to blood flow, retrievability, and interventionalist’s preference determine which filter is placed.

The permanent filters currently on the market in the United States include: Greenfield stainless steel and titanium (Boston Scientific/Meditech), Bird’s Nest (Cook), Simon Nitinol (Bard), VenaTech LP and LGM (B. Braun), and TrapEase (Cordis). The optional filters include: Gunther Tulip (Cook), OptEase (Cordis), and G2 (Bard). The only temporary filter available is the Tempo Filter II (B. Braun). There are many other filters not mentioned here available outside the United States, in clinical trials, or in premarketing.

The TrapEase filter is a double-basket, symmetrical device made from a single nitinol tube. The device has a non-expanded maximum length of 65 mm, which reduces to 50 mm when expanded to its maximum diameter of 35 mm; six straight struts connect the proximal and distal baskets. The proximal and distal basket is composed of six petal-shaped openings, which are connected by six straight struts. A proximal and distal hook is located on each connecting strut for fixation of the filter to the vena cava wall [5].

Complications of IVC filters can occur during placement, immediately following the procedure, or months to years later [7]. Hann et al. describes procedure-related complications including misplacement (1.3% of insertions), pneumothorax (0.02%), hematoma (0.6%), air embolism (0.02%), carotid artery puncture (0.04%), and arteriovenous fistula (0.02%). Fatalities caused by these complications are rare; they occur in only 0.13% of insertions. A common early post-procedure complication of filter placement is insertion site thrombosis (0.4-36%). Delayed complications of filter placement include: IVC thrombosis (2-10%), filter migration (0.3%) [8], IVC penetration (0.3%) [9], and filter disruption from entrapment of guide wires [1].

Migration is a rare complication of IVC filter placement. It is defined as shifting of the IVC filter more than 1 cm in a cranial or caudal direction [9]. Migration of the filter to the heart or pulmonary arteries is uncommon.

Signs and symptoms of IVC filter migration to the heart include: chest pain, hypotension, dyspnea, tachycardia, premature ventricular contractions, right bundle branch block, asystole, neck pain, atrial fibrillation, confusion, lightheadedness, diaphoresis, and nausea. The patient in this case presented with symptoms of confusion and lethargy for approximately one week. Two cases have been reported with asymptomatic patients. In those cases, one filter was located in the right atrium and the other was in the right pulmonary artery [10, 11].

Causes of filter migration can be grouped into three categories: mechanical, iatrogenic, and physiologic. Mechanical causes include failure of the device delivery system or the filter itself. These occurrences are usually realized during the implantation procedure and should be reported to the FDA and the device manufacturer.

Iatrogenic causes of filter migration include entanglement with guide wires for central venous line placement and attempting to place a filter without performing an inferior vena cavogram to properly size the device. Each filter is approved for deployment up to a specific inferior vena cava diameter. For example, the TrapEase filter is approved for inferior vena cava diameters 18-30 mm [5].

Physiologic causes of migration may result from temporary dysmorphism of the inferior vena cava. Bending, coughing, or Valsalva maneuvers may allow dislodgement of the filter. A “sail effect” in which blood flow might induce cephalad migration of conical filters was proposed by Rossi et al [12]. The force of a large thrombus striking the filter may be enough to cause dislodgement. Finally, a filter which already is bearing a burden of clot in addition to an insult listed above may synergistically induce migration.

The TrapEase IVC filter has been the subject of four clinical studies, two prospective, and two retrospective from 2001 to 2006. Among the 1,047 subjects, there were no cases of filter migration [5, 13-15]. Only two previous case reports of TrapEase filter migration to the heart have been published in the literature. In one case the patient died of a sudden cardiac arrest three days after filter deployment. The filter was found within the tricuspid valve on autopsy [3]. In the second case, the patient presented with dyspnea, hypotension, and confusion seven days following the filter placement. The filter was surgically removed and the patient was discharged on oral anticoagulation [4].

Conclusion

The number of IVC filters placed has increased from 2,000 in 1979 to 49,000 in 1999 [16]. In the literature, there are two reported cases of the TrapEase filter migration to the heart or pulmonary outflow tract. Our case is only the third occurrence; therefore, TrapEase IVC filter migration to the heart after successful placement is an extremely rare complication.

From our literature review of migrated and misdeployed filters, initial percutaneous intervention was successful in twenty-four out of thirty-one (77%) cases. When the filter could not be retrieved percutaneously, surgical excision was successful in three out of four (75%) cases, and the no further interventions group had four patients out of five (80%) survive at six, thirty-six, sixty, and eighty months follow-up respectively. Primary surgical intervention has been successful in fourteen out of nineteen (74%) cases since 1976, including our case. Six of the fifteen (40%) patients that did not undergo any intervention survived nine to forty-eight months.

Based on the data collected, it is difficult to make treatment recommendations. Percutaneous and surgical interventions have similar survival. At our institution, percutaneous retrieval is preferred, when possible, to minimize peri-operative morbidity. If the patient is too high risk for surgery and percutaneous intervention is not possible or unsuccessful, leaving the filter in place may be considered. In that situation, anticoagulation and/or a second filter are advisable.

Conflict of Interest

There is no grant support or conflict of interest to report by any of the authors.


References▴Top 
  1. Hann CL, Streiff MB. The role of vena caval filters in the management of venous thromboembolism. Blood Rev. 2005;19(4):179-202.
    pubmed doi
  2. Decousus H, Leizorovicz A, Parent F, Page Y, Tardy B, Girard P, Laporte S, et al. A clinical trial of vena caval filters in the prevention of pulmonary embolism in patients with proximal deep-vein thrombosis. Prevention du Risque d'Embolie Pulmonaire par Interruption Cave Study Group. N Engl J Med. 1998;338(7):409-415.
    pubmed doi
  3. Haddadian B, Shaikh F, Djelmami-Hani M, Shalev Y. Sudden cardiac death caused by migration of a TrapEase inferior vena cava filter: case report and review of the literature. Clin Cardiol. 2008;31(2):84-87.
    pubmed doi
  4. Porcellini M, Stassano P, Musumeci A, Bracale G. Intracardiac migration of nitinol TrapEase vena cava filter and paradoxical embolism. Eur J Cardiothorac Surg. 2002;22(3):460-461.
    pubmed
  5. Rousseau H, Perreault P, Otal P, Stockx L, Golzarian J, Oliva V, Reynaud P, et al. The 6-F nitinol TrapEase inferior vena cava filter: results of a prospective multicenter trial. J Vasc Interv Radiol. 2001;12(3):299-304.
    pubmed doi
  6. Goldfaber SZ, Grassi CJ. Management of pulmonary embolism. In: Sabeston DC, ed. Textbook of surgery, 8th ed. Philadelphia: WB Saunders, 1990;115-127.
    pubmed
  7. Ray CE, Jr., Kaufman JA. Complications of inferior vena cava filters. Abdom Imaging. 1996;21(4):368-374.
    pubmed doi
  8. Streiff MB. Vena caval filters: a review for intensive care specialists. J Intensive Care Med. 2003;18(2):59-79.
    pubmed doi
  9. Mitchell WB, Bonn J. Percutaneous retrieval of a Greenfield filter after migration to the left pulmonary artery. J Vasc Interv Radiol. 2005;16(7):1013-1017.
    pubmed
  10. Patel A, Cholankeril J. Case of the migrating embolic filter. Hosp Pract. 1991;26(4):129-132.
    pubmed
  11. Rao KM, Simons AJ, Hare CL, Smulyan H. Migration of a Kimray-Greenfield filter into the pulmonary artery: localization by transesophageal echocardiography. Am Heart J. 1993;125(2 Pt 1):543-544.
    pubmed doi
  12. Rossi P, Arata FM, Bonaiuti P, Pedicini V. Fatal outcome in atrial migration of the Tempofilter. Cardiovasc Intervent Radiol. 1999;22(3):227-231.
    pubmed doi
  13. Schutzer R, Ascher E, Hingorani A, Jacob T, Kallakuri S. Preliminary results of the new 6F TrapEase inferior vena cava filter. Ann Vasc Surg. 2003;17(1):103-106.
    pubmed doi
  14. Kalva SP, Wicky S, Waltman AC, Athanasoulis CA. TrapEase vena cava filter: experience in 751 patients. J Endovasc Ther. 2006;13(3):365-372.
    pubmed
  15. Liu WC, Do YS, Choo SW, Kim DI, Kim YW, Kim DK, Shin SW, et al. The mid-term efficacy and safety of a permanent nitinol IVC filter(TrapEase). Korean J Radiol. 2005;6(2):110-116.
    pubmed
  16. Stein PD, Kayali F, Olson RE. Twenty-one-year trends in the use of inferior vena cava filters. Arch Intern Med. 2004;164(14):1541-1545.
    pubmed doi


This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


Journal of Medical Cases is published by Elmer Press Inc.

 

Browse  Journals  

 

Journal of Clinical Medicine Research

Journal of Endocrinology and Metabolism

Journal of Clinical Gynecology and Obstetrics

 

World Journal of Oncology

Gastroenterology Research

Journal of Hematology

 

Journal of Medical Cases

Journal of Current Surgery

Clinical Infection and Immunity

 

Cardiology Research

World Journal of Nephrology and Urology

Cellular and Molecular Medicine Research

 

Journal of Neurology Research

International Journal of Clinical Pediatrics

 

 
       
 

Journal of Medical Cases, monthly, ISSN 1923-4155 (print), 1923-4163 (online), published by Elmer Press Inc.                     
The content of this site is intended for health care professionals.
This is an open-access journal distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International License, which permits unrestricted
non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Creative Commons Attribution license (Attribution-NonCommercial 4.0 International CC-BY-NC 4.0)


This journal follows the International Committee of Medical Journal Editors (ICMJE) recommendations for manuscripts submitted to biomedical journals,
the Committee on Publication Ethics (COPE) guidelines, and the Principles of Transparency and Best Practice in Scholarly Publishing.

website: www.journalmc.org   editorial contact: editor@journalmc.org    elmer.editorial@hotmail.com
Address: 9225 Leslie Street, Suite 201, Richmond Hill, Ontario, L4B 3H6, Canada

© Elmer Press Inc. All Rights Reserved.


Disclaimer: The views and opinions expressed in the published articles are those of the authors and do not necessarily reflect the views or opinions of the editors and Elmer Press Inc. This website is provided for medical research and informational purposes only and does not constitute any medical advice or professional services. The information provided in this journal should not be used for diagnosis and treatment, those seeking medical advice should always consult with a licensed physician.