Case Report
Volume 6, Number 2, February 2015, pages 98-100
An Unusual Complication of Intraosseous Needle Use
John T. Dennya, c, Andres Ocampoa, Angela M. Dennyb, Wayne Wanga, Marco Del Castilloa, Julia E. Dennyb, Darrick Chyua, Geza Kissa
aDepartment of Anesthesia, Rutgers Robert Wood Johnson Medical School, Suite 3100 CAB, 125 Paterson Street, New Brunswick, NJ 08901, USA
bRutgers Graduate School of Nursing, 65 Bergen Street, Newark, NJ 07107, USA
cCorresponding Author: John T. Denny, Department of Anesthesia, Rutgers Robert Wood Johnson Medical School, Suite 3100 CAB, 125 Paterson Street, New Brunswick, NJ 08901, USA
Manuscript accepted for publication January 07, 2015
Short title: Complication of Intraosseous Needle Use
doi: http://dx.doi.org/10.14740/jmc2080e
Abstract | ▴Top |
Intraosseous (IO) needles are extremely useful tools to gain urgent access to the circulation when more conventional routes are unavailable or unsuccessful. As with any medical procedure, complications can occur. We review the indications for IO usage and illustrate an unusual complication of continued serous drainage following IO needle removal. The drainage was significant enough to require use of a drainage bag. This chronically ill patient developed septic shock, multi-system organ failure, renal failure and a high catabolic state. As he did not fully tolerate enteric tube feeds, it was not possible to maintain adequate serum albumin. The resultant lower colloid oncotic pressure contributed to the chronic serous leak from the IO site. Avoidance of such a complication relies on proper IO placement and the ability to maintain adequate nutrition to maintain plasma oncotic pressure. Other complications of IO use are reviewed.
Keywords: Intraosseous needle; Complications of intraosseous use; Intravenous access
Introduction | ▴Top |
Intraosseous (IO) needles were developed for urgent circumstances where intravenous (IV) access is needed but may be difficult to achieve due to a variety of factors. Obtaining IV access in the pre-hospital setting can often be a problem where environmental factors include cramped quarters, poor lighting or a moving platform of transportation [1, 2]. In both pre-hospital and hospital locations, individual patient factors can also contribute, such as poor venous access or anatomy, a history of IV drug use or morbid obesity. Clinical factors such as peripheral vasoconstriction in patients in shock or in hypothermic patients, or burn victims can also make IV access challenging. Victims of major trauma, out-of-hospital cardiac arrest or sepsis may require immediate vascular access.
Drinker et al first described IO infusion in 1922 [3]. In this vascular access technique, the marrow of a mammalian bone becomes a “non-collapsible vein” into which fluids can be infused. Although it became popular in the United States in the 1940s for battlefield resuscitation [4], its use declined with the improvement in IV devices and increased practitioner IV skill [5]. IO was rediscovered in the 1980s and is now the preferred avenue of access when a peripheral IV cannot be readily established [6]. It is a reliable method for administering drugs, fluid and blood. Complications are rare and most often not significant.
Case Report | ▴Top |
We report a case where appropriate IO use unfortunately subsequently led to a chronic serous leak from the puncture site in the patient’s right tibia. Patient was a 67-year-old male with a past medical history of hypertension, adult onset diabetes mellitus, hyperlipidemia and end stage renal disease. He was status post a deceased donor renal transplant 4 months previously. One month post-operatively, he was admitted for a right common femoral DVT and placed on coumadin. He was later followed up in clinic. On this admission, patient presented with an elevated creatinine, hydronephrosis, allograft dysfunction and a supratherapeutic international normalized ratio of 12. After his coagulopathy was partially corrected, he underwent a cystoscopy. Urine and blood cultures grew coagulase negative Staphylococcus. Eleven days after admission, patient became hypotensive with a systolic blood pressure in the 80s, and a sinus tachycardia of 140. He also complained of shortness of breath, and pulse oximetry saturation decreased. He was endotracheally intubated with cricoid pressure; however, feculent fluid was noted in the oropharynx and likely aspirated. Attempts were made to gain central venous access, but were initially unsuccessful. Thus an IO needle was urgently placed by an experienced operator in the right tibia for resuscitation. Broad spectrum antibiotics were started, immunosuppression medications were held and stress dose steroids were given. In the subsequent hospital course he developed septic shock, and was treated with vasopressors, fluid resuscitation, mechanical ventilation and antibiotics. Cultures grew out resistant Enterobacter cloacae. Patient developed renal failure and was placed on continuous veno-venous hemofiltration. He developed continuous serous drainage from the IO site after the needle was removed. This clear straw-colored fluid was collected using a drainage bag (Fig. 1).
Figure 1. Drainage bag over previous IO needle site. |
Patient’s overall predicted mortality risk was > 95% using Acute Physiology and Chronic Health Evaluation (APACHE) scoring. He went on to develop worsening multi-system organ failure (MSOF). Ultimately, his family withdrew support and patient died.
Discussion | ▴Top |
This is a novel complication of IO use involving chronic serous drainage from the puncture site. Avoidance of such a complication relies on proper IO placement and the ability to maintain adequate nutrition to maintain plasma oncotic pressure. This chronically ill patient developed septic shock, MSOF, renal failure and a high catabolic state. As he did not fully tolerate enteric tube feeds, it was not possible to maintain adequate serum albumin. The resultant lower colloid oncotic pressure contributed to the chronic serous leak from the IO site.
Many different complications are reported with IO use. Breaking the IO needle was described by Danz et al [7]. Fractures of the bone pierced by the needle have been reported [8, 9]. Bilateral tibial fractures have also been reported [10]. Even sternal perforation and hemomediastinum have been reported after IO placement in the sternum [11]. However, the most common complication of IO use is extravasation. Simmons et al report two cases of extravasation, one leading to a compartment syndrome [12]. If IO extravasation is more severe, compartment syndrome has been described [13]. Unrecognized compartment syndrome has led to eventual amputation [14].
Conclusion
In this report, we reviewed the indications for IO usage. Other complications of IO use are reviewed. We report a novel complication of IO use involving chronic serous drainage from the puncture site. Avoidance of such a complication relies on proper IO placement and the ability to provide adequate nutrition to maintain plasma oncotic pressure.
Conflict of Interest
Authors have no disclosures or conflicts of interest.
References | ▴Top |
- Sunde GA, Heradstveit BE, Vikenes BH, Heltne JK. Emergency intraosseous access in a helicopter emergency medical service: a retrospective study. Scand J Trauma Resusc Emerg Med. 2010;18:52.
doi pubmed - Lyon RM, Donald M. Intraosseous access in the prehospital setting-ideal first-line option or best bailout? Resuscitation. 2013;84(4):405-406.
doi pubmed - Drinker CK, Drinker KR, Lund CC. The circulation in the mammalian bone-marrow. Am J Physiol. 1922;62.
- Bailey H. Bone-marrow Injections. Br Med J. 1944;1(4335):181-182.
doi pubmed - Luck RP, Haines C, Mull CC. Intraosseous access. J Emerg Med. 2010;39(4):468-475.
doi pubmed - Phillips B, Zideman D, Garcia-Castrillo L, Felix M, Shwarz-Schwierin V. European Resuscitation Council Guidelines 2000 for Advanced Paediatric Life Support. A statement from Paediatric Life Support Working Group and approved by the Executive Committee of the European Resuscitation Council. Resuscitation. 2001;48(3):231-234.
doi - Danz M, Schulz G, Hinkelbein J, Braunecker S. Breaking the needle: a rare complication on EZ-IO removal. Eur J Anaesthesiol. 2014;31(3):176-177.
doi pubmed - Bowley DM, Loveland J, Pitcher GJ. Tibial fracture as a complication of intraosseous infusion during pediatric resuscitation. J Trauma. 2003;55(4):786-787.
doi pubmed - Katz DS, Wojtowycz AR. Tibial fracture: a complication of intraosseous infusion. Am J Emerg Med. 1994;12(2):258-259.
doi - La Fleche FR, Slepin MJ, Vargas J, Milzman DP. Iatrogenic bilateral tibial fractures after intraosseous infusion attempts in a 3-month-old infant. Ann Emerg Med. 1989;18(10):1099-1101.
doi - Plancade D, Millot I, Fetissof H, Landy C, Schaeffer E, Perez JP, Nadaud J. Sternal perforation with an intraosseous device and hemomediastinum infusion. Ann Fr Anesth Reanim. 2013;32(3):e69-70.
doi pubmed - Simmons CM, Johnson NE, Perkin RM, van Stralen D. Intraosseous extravasation complication reports. Ann Emerg Med. 1994;23(2):363-366.
doi - Vidal R, Kissoon N, Gayle M. Compartment syndrome following intraosseous infusion. Pediatrics. 1993;91(6):1201-1202.
pubmed - Moscati R, Moore GP. Compartment syndrome with resultant amputation following intraosseous infusion. Am J Emerg Med. 1990;8(5):470-471.
doi
This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Journal of Medical Cases is published by Elmer Press Inc.